Error Information
General Error Messages

Fan Failed

Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the main board to monitor the fan
speed. If the fan speed falls below a certain limit for a certain length of time,
the error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are
turned off to assure that the module does not overheat inside.
Probable cause
1
2
3

Leak

Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the main board.
Probable cause
1
2
3
4
100
Fan cable disconnected.
Defective fan.
Defective main board.
Loose fittings.
Broken capillary.
Loose or leaking purge valve, inlet
valve, or outlet valve.
Defective pump seals.
Suggested actions
Please contact your Agilent service representative.
Please contact your Agilent service representative.
Please contact your Agilent service representative.
Suggested actions
Ensure all fittings are tight.
Exchange defective capillaries.
Ensure pump components are seated correctly. If
there are still signs of a leak, exchange the
appropriate seal (purge valve, inlet valve, outlet valve).
Exchange the pump seals.