4.2.4 Procedure while error 38 – Lid motor is blocked
The lid motor is blocked and for 10 seconds the error massage "error 38" is shown in the display
"time" (A-3). After expiring of this error message please press the key "lid" (7). If the lid lock doesn´t
loose, repeat this procedure two or three more times. If the lid lock still doesn`t loose, the lid must be
opened by using the manual emergency release. Proceed as described in chapter 4.2.1. After
opening the centrifuge lid, switch on again the device. The lid lock adjusts itself in the basic position.
The closing process can be carried out normally again.
If these errors occurs again, please contact the service.

5. RECEIPT OF CENTRIFUGES FOR REPAIR

Health risk from contaminated equipment, rotors and accessories
In case of returning the centrifuge for repairing, to the manufacturer, please know the following:
The centrifuge must be decontaminated and cleaned, before shipment, for the protection of
persons, environment and material.
Decontamination certificate at goods return delivery, (see APPENDIX P. XIII)
We reserve the right to accept contaminated centrifuges.
Furthermore, all costs that may have occurred during the cleaning and disinfection of the units, will
go to the debit of the customer's account.
We ask for your understanding.
Return of Power Cords
In case of a return of a centrifuge, we also ask you to send its power cord. This eliminates the risk of
a faulty power cord. If no power cord is attached to the centrifuge, a new one will be delivered and
charged.
We ask for your understanding.
36
Z366K_V1.22_eng
© Hermle Labortechnik GmbH

Advertisement

Table of Contents
loading

Table of Contents