Type BF Patient Leakage Current Test - Probes (This test should be done for all probe)
Tester
Tester Polarity
Neutral
Switch
Closed
Normal
Closed
Normal
Closed
Reversed
Closed
Reversed
Open
Normal
Open
Reversed
Type CF Patient Auxiliary Current Test - ECG Lead to Probes
Tester
Tester Polarity
Neutral
Switch
Closed
Normal
Closed
Normal
Closed
Reversed
Closed
Reversed
Open
Normal
Open
Reversed
Type BF Patient Leakage Current Test - Mains to Probe
Probe (This test should be done for all probes)
Tester
Tester Polarity
Neutral
Switch
Closed
Normal
Closed
Reversed
Type CF Patient Leakage Current Test - Mains to ECG leads
ECG leads connected all together
Tester
Tester Polarity
Neutral
Switch
Closed
Normal
Closed
Reversed
6
DRAFT (J
EVISION
Ground Switch
Closed
Open
Closed
Open
Closed
Closed
Ground Switch
Closed
Open
Closed
Open
Closed
Closed
Ground Switch
Closed
Closed
Ground Switch
Closed
Closed
Chapter 10 - Care & Maintenance
17, 2013)
ANUARY
Max Value
Value
Allowed
Measured
0.1mA
0.5mA
0.1mA
0.5mA
0.5mA
0.5mA
Max Value
Value
Allowed
Measured
0.01mA
0.05mA
0.01mA
0.05mA
0.05mA
0.05mA
Max Value
Value
Allowed
Measured
5mA
5mA
Max Value
Value
Allowed
Measured
0.05mA
0.05mA
V
® P8 / V
OLUSON
OLUSON
P
S
ROPRIETARY
ERVICE
OK?
Comments
OK?
Comments
OK?
Comments
OK?
Comments
® P6
M
ANUAL
10-35