7-4-1-1
How to use a Capture Filter
Use filters to limit the captured data. Filter
by IP and port number
1.) Click on CAPTURE FILTER (see:
2.) Select NEW.
3.) Edit "Filter name". For example: DICOM port 104
4.) Enter the "Filter string" with the following syntax: port xxx and host yy.yy.yy.yy,
where xxx is the number of "My computer" in the scanner and yy.yy.yy.yy is the IP address of the
device under test (PACS, server. etc.)
5.) Click OK.
by IP only
1.) Click on CAPTURE FILTER (see:
2.) Select IP only.
3.) Use the IP address of the scanner or the one from the DICOM device under test on the "Filter string"
field, using the following syntax: host yy.yy.yy.yy (example: host 3.62.21.12).
4.) Click OK.
6
DRAFT (J
EVISION
Figure 7-8 Capture Filter
Chapter 7 - Diagnostics/Troubleshooting
17, 2013)
ANUARY
by IP and port number
Figure 7-9 on page
7-9).
Figure 7-9 on page
7-9).
V
® P8 / V
OLUSON
OLUSON
P
S
ROPRIETARY
ERVICE
or
by IP
only.
® P6
M
ANUAL
7-9