GE Voluson P6 Service Manual page 263

Hide thumbs Also See for Voluson P6:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
DIRECTION 5459672-100, R
5-14-1-3
Rollback
This function offers the possibility to simply restore the previously used system configuration (rollback),
which was stored on R:\ during "FMI from DVD".
1.) Turn system OFF and then back ON.
2.) As soon as the "Boot Screen" appears (see:
[↓ Arrow down] key on the keyboard until the Rollback item is highlighted, then press ENTER.
3.) When the following WARNING message appears, press the ← [Arrow left] button to highlight OK
and then press ENTER.
select [OK]
After performing the rollback, the system reboots.
The Ultrasound Application is started and finally the 2D screen is displayed on the monitor.
5-14-1-4
Memtest
Starts the PC Memory Test.
Operation see:
5-56
6
DRAFT (J
EVISION
Figure 5-37 Warning message
Section 5-15-8 "Memory Check in LINUX" on page
Section 5-14 - Boot Screen Functions
17, 2013)
ANUARY
Figure 5-36 on page
V
® P8 / V
OLUSON
OLUSON
P
S
ROPRIETARY
ERVICE
5-55), press the PG DN
5-77.
® P6
M
ANUAL