GE Voluson P6 Service Manual page 522

Hide thumbs Also See for Voluson P6:
Table of Contents

Advertisement

GE H
DRAFT
EALTHCARE
DIRECTION 5459672-100, R
10-6-5
Physical Inspection
Table 10-12 Physical Checks
Step
1
Labeling
2
Scratches & Dents
3
LCD Monitor Display
Control Panel and
4
Keyboard
5
DVD+R/RW Drive
6
Wheels & Brakes
7
Cables & Connectors
8
Power Cord
9
Shielding & Covers
10
Peripherals
11
External I/O
12
Op Panel Lights
10-6-6
Optional Diagnostic Checks
Optionally you can access the diagnostic software as described in Chapters 5 or 7.
View the error logs and run desired diagnostics.
6
DRAFT (J
EVISION
Item
Verify that all system labeling is present and in readable condition.
Inspect the console for dents, scratches or cracks.
Inspect the LCD Monitor Display for scratches and raster burns.
Verify proper operation of Contrast and Brightness controls.
Inspect the Control Panel and Keyboard. Note any damaged or missing items.
(Replace faulty components, as required).
Verify proper operation of Control Panel backlighting and TGC sliders.
Clean the drive head and media with the vendor-supplied cleaning kit.Advise the user to repeat
this often, to prevent future problems. DVDs/CDs must be stored away from dust and cigarette
smoke. Do not use alcohol or benzene to clean the drive
Check all wheels and casters for wear and verify operation of foot brake, to stop the unit from
moving, and release mechanism. Check all wheel locks and swivel locks for proper operation.
Check all internal cable harnesses and connectors for wear and secure connector seating.
Pay special attention to probe strain or bend reliefs.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation or other
deterioration, and verify continuity. Tighten the clamps that secure the power cord to the unit and
the outlet plug to the cord.
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Missing covers and hardware could cause EMI/RFI problems during scanning.
Check and clean the peripherals according to the manufacturer's directions.
To prevent EMI or system overheating, dress the peripheral cables inside the peripheral cover.
Check all connectors for damage and verify that the labeling is good.
Check proper operation of all control panel key illuminations (flash once during system start-up).
Chapter 10 - Care & Maintenance
17, 2013)
ANUARY
Description
V
® P8 / V
OLUSON
OLUSON
P
S
M
ROPRIETARY
ERVICE
® P6
ANUAL
10-11

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voluson p8