GE H
RAFT
EALTHCARE
DIRECTION 5459672-100, R
Section 8-41
Replacing optional Peripherals
NOTICE
Normally auxiliary devices and peripherals come pre-installed with the Voluson® P8 / Voluson® P6
system.
!! NOTICE:
8-41-1
Manpower
One person, 70 minutes
8-41-2
Tools
Standard Field Service Tool Kit
NOTICE
READ and UNDERSTAND these instructions thoroughly before proceeding with the installation.
Perform each step in sequence and check it off when completed. If a problem occurs after completion,
!! NOTICE:
go back through the procedure and check for implementation errors before contacting your Ultrasound
Region.
Possible Operational damage. Failure to strictly follow ESD (Electrostatic Discharge)
CAUTION
precautions during this installation may cause constant or intermittent operational
!! CAUTION:
abnormalities. Strictly follow all precautions.
8-41-3
Printer Installation (BW and Color)
8-41-3-1
BW Printer: SONY UP-D897, Color Printer: SONY UP-D25MD
1.) Assemble 2 EMI gaskets (5448934 - EMI gasket, side, b-w printer bkt-3) on printer base plate.
2.) Assemble 2 EMI gaskets (5447303 - EMI gasket, side, b-w printer bkt-2) on printer base plate.
3.) Assemble 2 EMI gaskets (5457475 - EMI gasket, side, b-w printer bkt-1) on printer base plate.
4.) Assemble printer base pate with B/W printer using 8 screws (2306562-screw FH M3 x 6).
8-88