2.4. Measurement or input functions
20
Select the type of the signal to be measured by using the menus and use the corresponding terminals:
IN
a)
Press ENTER to select volt measurement; press , , and ➔ to select another signal.
The other magnitudes follow the same selection process.
When performing the measurement of OHM, you must also select the 2-, 3- or 4-wire options. For TC
(thermocouple), you must select the thermocouple type and cold junction compensation type: Internal or Manual.
With the Internal option, compensation is performed internally; in Manual, it is necessary to provide the calibrator
with the cold junction temperature value, by entering the digits through the numeric keypad.
Selects the input function and the ENTER key should be pressed.
V
F
IN = x.xxxx V
C/CE
mV
mA
Ohm
TC
RTD
SW
Display indicates volt input.
Goes back to previous menu
OP
NO