Replacing the
UV Lamp
DOC-24333 Rev. A
The UV lamp used for curing models has a long, but limited, working life.
The Stratasys service engineer tests its effectiveness during regular
maintenance checks, and replaces it, if necessary. If, for any reason, you
need to replace the UV lamp, follow these instructions:
1. Start the UV Lamp Replacement Wizard from the Options menu (see
figure 6‐34 on page 28).
WARNING: Before continuing, make sure that the safety interlock in the
printer cover is not defeated, and that the lamp is not hot.
The online/offline button at the bottom of the printer interface should be
red..
Figure 6-58: Offline mode indicator (red)
2. Make sure that the build tray is empty, and confirm this in the wizard
screen. Then, close the printer cover and click
Figure 6-59: "Empty tray" confirmation screen
The axes move to the "home" position.
Objet30 User Guide
Next
.
6–39