Protective Ground/Earth Measurement - Siemens MOBILETT Plus HP Maintenance Instruction

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

Inspection and Maintenance
2.17

Protective ground/earth measurement

Perform the protective conductor test, with all system covers closed, according to
ARTD-002.731.17... (safety-related regulations on installation and maintenance). The
measured values are to be entered in the maintenance protocol. The protective con-
ductor resistance must not exceed 0.2 Ohms.
SIE
Measured value
Siemens AG
Medical Solutions
SPR8-220.101.04.03.02
07.05
Page 59 of 64
CS PS 24
59
MOBILETT Plus HP