32
2.2.8
Arm system and single tank
Fig. 10:
SIM
Movement of the arm system
SIM
Movement of the single tank fork
SIM
Single tank movement
•
The entire arm system including the single tank must not move by itself in any position.
However, it must be easily moveable to a new position. If this is not the case, proceed
according to the section "Adjusting the friction linings" in the Service Instructions.
•
Check that the single tank fork can be easily turned and stays in any desired position.
•
Check that the single tank moves easily in the fork and stays in any desired position.
MOBILETT Plus HP
SPR8-220.101.04.03.02
07.05
Inspection and Maintenance
Page 32 of 64
CS PS 24
Siemens AG
Medical Solutions