Inspection and Maintenance
Fig. 26:
Fig. 27:
•
Calculate the coefficient:
Fig. 28:
In this example, C=0.0105 and is therefore ≤ 0.0450,
⇒
i.e., the generator is within specifications.
•
Reattach the upper cover.
Siemens AG
Medical Solutions
SPR8-220.101.04.03.02
07.05
Page 49 of 64
CS PS 24
49
MOBILETT Plus HP