MAMMOMAT 3000
Service
Repair Instructions
Replacing wires in stereotactic device
Register 5
Print No.: SPB7-230.091.09.02.02
Replaces: N/A
All manuals and user guides at all-guides.com
SP
© Sie mens AG 1999
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 01.99
65 19 941

Advertisement

loading

Summary of Contents for Siemens MAMMOMAT 3000

  • Page 1 All manuals and user guides at all-guides.com MAMMOMAT 3000 Service Repair Instructions Replacing wires in stereotactic device © Sie mens AG 1999 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages.
  • Page 2 All manuals and user guides at all-guides.com 0 - 2 Revision Chapter Page Revision MAMMOMAT 3000 Register 5 SPB7-230.091.09 Page 2 of 4 Siemens-Elema AB Service Rev. 02 01.99 SPS-UD Solna, Sweden...
  • Page 3: Table Of Contents

    Reassembly ........2 - 3 Siemens-Elema AB Register 5 SPB7-230.091.09 Page 3 of 4 MAMMOMAT 3000 Solna, Sweden Rev. 02 01.99 SPS-UD...
  • Page 4 All manuals and user guides at all-guides.com 0 - 4 Contents This page intentionally left blank. MAMMOMAT 3000 Register 5 SPB7-230.091.09 Page 4 of 4 Siemens-Elema AB Service Rev. 02 01.99 SPS-UD Solna, Sweden...
  • Page 5: Purpose

    The pozidrive screws can be very hard to loosen. When unscrewing, be sure of enough grip with the pozidrive screw-driver, so that the screws do not get damaged. Siemens-Elema AB Register 5 SPB7-230.091.09 Page 1 of 2 MAMMOMAT 3000 Solna, Sweden Rev. 02 01.99 SPS-UD Service...
  • Page 6 All manuals and user guides at all-guides.com 1 - 2 Prerequisites This page intentionally left blank. MAMMOMAT 3000 Register 5 SPB7-230.091.09 Page 2 of 2 Siemens-Elema AB Service Rev. 02 01.99 SPS-UD Solna, Sweden...
  • Page 7: Dismounting

    - Two screws at either side of the biopsy unit. - Four screws at the bottom of the biopsy unit. Place the biopsy unit as shown in figure 1 (pos. 4) Siemens-Elema AB Register 5 SPB7-230.091.09 Page 1 of 4 MAMMOMAT 3000 Solna, Sweden Rev. 02 01.99 SPS-UD Service...
  • Page 8 Dismounting of wires Disconnect the grounding wire by carefully unscrewing the screw. Remove the locking washers one on each side. Unhook the springs. Remove the wires with handels. MAMMOMAT 3000 Register 5 SPB7-230.091.09 Page 2 of 4 Siemens-Elema AB Service Rev.
  • Page 9: Reassembly

    Secure the ring and the spring with the locking washer. Reassemble the biopsy unit. NOTICE Don´t forget to connect the grounding wire. Siemens-Elema AB Register 5 SPB7-230.091.09 Page 3 of 4 MAMMOMAT 3000 Solna, Sweden Rev. 02 01.99 SPS-UD Service...
  • Page 10 All manuals and user guides at all-guides.com 2 - 4 Procedure This page intentionally left blank. MAMMOMAT 3000 Register 5 SPB7-230.091.09 Page 4 of 4 Siemens-Elema AB Service Rev. 02 01.99 SPS-UD Solna, Sweden...