Inspection and Maintenance
2.11

DAP measuring system (optional)

Press the test button on the display. The value on the display should be in the range
80-120 µGym
high sensitivity, i.e., DAP resolution 0.01 µGym
to the Service Instructions.
X
Check the function of the DAP measuring system by releasing an exposure. Check if a
measured value is shown on the display.
SIE
Function of DAP measuring system
Siemens AG
Medical Solutions
2
(DAP resolution 0.1 µGym
SPR8-220.101.04.03.02
07.05
2
2
) or 8-12 µGym
(when using a chamber of
2
). If not, calibrate the system according
Page 51 of 64
CS PS 24
51
MOBILETT Plus HP