54
2.13
Checking the luminance
Fig. 32:
Required measurement devices: lux meter
NOTE
PMF
Luminance
If the collimator lamp is switched on for the purpose of illuminating the radiation field, the
average luminance must be 180 lux at a distance of 100 cm. The average luminance is
determined by taking measurements in the approximate center of each of the four quad-
rants of the light
MOBILETT Plus HP
If the MOBILETT is equipped with a DAP measuring system, re-
move the DAP ionization chamber first.
field,(Fig. 32 / p. 54)
SPR8-220.101.04.03.02
07.05
Inspection and Maintenance
.
Page 54 of 64
CS PS 24
Siemens AG