Siemens MOBILETT Plus HP Maintenance Instruction page 16

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

16
Fig. 6:
Pos. 1
Pos. 2
Pos. 3
Pos. 4
WARNING
MOBILETT Plus HP
Left connector on D7 to the + connector of the digital multimeter
Second left connector on D7 to the fuse side (U11) next to the capacitor
Second right connector on D7 to the - connector (GND) of the digital multimeter
Right connector on D7 to the P11 connector on the power conversion unit (M10)
If the charging/discharging does not function, the power conver-
sion unit M10 and M11 must be considered "charged".
This means a risk of high voltage.
Use caution when performing measurements at the
capacitors in M10 and M11. Use only the specified mea-
surement devices (350 V DC).
SPR8-220.101.04.03.02
07.05
Page 16 of 64
CS PS 24
General
Siemens AG
Medical Solutions