52
2.12
Replacing the collimator lamp
Fig. 29:
Fig. 31:
PMP
Replace every year.
•
The collimator lamp must be replaced every year.
•
Remove the cover
(Fig. 29 / p. 52)
•
Remove the protective plate
screws marked "B" by one turn. Do not remove them. Do not remove nuts "C", "D" and
"E" in this process
•
Remove the lamp and replace it with a new lamp,
glass bulb with your fingers.
•
Reinstall the protective plate and tighten the screws marked "B".
•
Reattach the cover
PMF
Function of the collimator lamp
MOBILETT Plus HP
(1/Fig. 29 / p. 52)
. Turn the collimator so that the lamp housing is accessible.
(2/Fig. 30 / p. 52)
(Fig. 30 / p. 52)
(1/Fig. 29 / p. 52)
SPR8-220.101.04.03.02
07.05
Inspection and Maintenance
Fig. 30:
by removing the four screws marked "A"
covering the collimator lamp. Loosen the
.
(Fig. 31 / p. 52)
.
Page 52 of 64
CS PS 24
. Do not touch the
Siemens AG
Medical Solutions