Siemens MOBILETT Plus HP Maintenance Instruction page 40

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

40
The "kV-" button allows you to run the test programs in reverse order. The "collimator
lamp" button is used for the ON/OFF function, i.e., it starts or stops the test program dis-
played.
Example of the display for P06 after activating the collimator lamp (starts the formation of
the capacitor bank):
kV
CAP
The test program is stopped by pressing the "collimator lamp" button a second time.
Example of the display for P06:
kV
P06
Program P06 is now stopped. A new service program can be selected with the "kV+/-"
buttons or program No. 6 can be restarted by pressing the "collimator lamp" button.
NOTE
MOBILETT Plus HP
mAs
For = capacitor formation program
mAs
Stop
Messages appearing in service mode, with the exception of "Quick
adjustment", are only displayed, i.e., they are not incorporated in
the error memory.
SPR8-220.101.04.03.02
07.05
Inspection and Maintenance
Page 40 of 64
CS PS 24
Siemens AG
Medical Solutions