44
kV Meter
selection
DC voltage
* The measurement inaccuracy of the respective measurement device must be sub-
tracted from these limit values.
Setting
•
Proceed according to section "P10 - Fast adaptation" in the Service Instructions in or-
der to adjust excessively high deviations.
PMF
kV accuracy 52 kV, 50 mAs (B)
PMF
kV accuracy 81 kV, 20 mAs (B)
PMF
kV accuracy 133 kV, 12.5 mAs (B)
•
Connect the power supply plug and switch the main switch to position "M"
•
X
Set the following exposure parameters and release an exposure:
kV Meter
selection
DC voltage
•
Disconnect the power supply plug, roll up the cable and switch the main switch to posi-
tion "B".
PMF
kV accuracy 81 kV, 20 mAs (M)
MOBILETT Plus HP
Exposure parameter
52 kV
81 kV
133 kV
Exposure parameter
81 kV
SPR8-220.101.04.03.02
07.05
Inspection and Maintenance
Limit values* to be maintained
50 mAs
20 mAs
12.5 mAs
Limit values* to be maintained
20 mAs
Page 44 of 64
CS PS 24
49.0 - 55.0 kV
77.0 - 85.0 kV
126.4 - 139.6 kV
77.0 - 85.0 kV
Siemens AG
Medical Solutions