Siemens MOBILETT Plus HP Maintenance Instruction page 44

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

44
kV Meter
selection
DC voltage
* The measurement inaccuracy of the respective measurement device must be sub-
tracted from these limit values.
Setting
Proceed according to section "P10 - Fast adaptation" in the Service Instructions in or-
der to adjust excessively high deviations.
PMF
kV accuracy 52 kV, 50 mAs (B)
PMF
kV accuracy 81 kV, 20 mAs (B)
PMF
kV accuracy 133 kV, 12.5 mAs (B)
Connect the power supply plug and switch the main switch to position "M"
X
Set the following exposure parameters and release an exposure:
kV Meter
selection
DC voltage
Disconnect the power supply plug, roll up the cable and switch the main switch to posi-
tion "B".
PMF
kV accuracy 81 kV, 20 mAs (M)
MOBILETT Plus HP
Exposure parameter
52 kV
81 kV
133 kV
Exposure parameter
81 kV
SPR8-220.101.04.03.02
07.05
Inspection and Maintenance
Limit values* to be maintained
50 mAs
20 mAs
12.5 mAs
Limit values* to be maintained
20 mAs
Page 44 of 64
CS PS 24
49.0 - 55.0 kV
77.0 - 85.0 kV
126.4 - 139.6 kV
77.0 - 85.0 kV