Inspection and Maintenance
2.4

Service mode

Fig. 20:
NOTE
Entering service mode
Switch S1:1 on board D1 must be switched over (S1:1 to ON) to call up the service pro-
gram. The program version is then displayed:
kV
vEr
(depends on the software version installed)
(To exit the service mode, switch S1:1 on board D1 must be switched back to OFF. This
will cause any program still running to be shut down.)
Selection of test programs with buttons "kV+/ -"
Several test programs are available. The respective program number appears in the kV
window of the display while the mAs window remains blank. Example of the display for
program No. 6:
kV
P06
By pressing the "kV+" button, the next test program is called up. As soon as the last test
program has been reached and "kV+" is pressed again, the display begins again with the
program version.
Siemens AG
Medical Solutions
Switch S10 on D7 OFF prior to the following inspection.
mAs
x.x = Version x.x
mAs
SPR8-220.101.04.03.02
07.05
Page 39 of 64
CS PS 24
39
MOBILETT Plus HP