Inspection and Maintenance
2.2.5
Transport safeguard
Fig. 8:
Pos. 1
Transport safeguard
SIM
Damage
SIM
Locking
•
Check the transport safeguard
age, the transport safeguard must be replaced.
•
Check that the handle of the arm system locks securely into the transport safeguard.
2.2.6
Handle
SIM
Mounting
•
Check that the handle on the support arm is securely mounted. If necessary, tighten the
screws with a torque of 15-20 Nm.
2.2.7
Collimator adjustment knobs
SIM
Mounting
•
Check whether the adjustment knobs on the collimator are tightened securely
(Fig. 9 / p.
Siemens AG
Medical Solutions
31). If necessary, tighten both Allen screws on each of the knobs.
SPR8-220.101.04.03.02
07.05
Fig. 9:
(Fig. 8 / p. 31)
for mechanical damage. In case of dam-
Page 31 of 64
CS PS 24
31
MOBILETT Plus HP