Inspection and Maintenance
2.2.5

Transport safeguard

Fig. 8:
Pos. 1
Transport safeguard
SIM
Damage
SIM
Locking
Check the transport safeguard
age, the transport safeguard must be replaced.
Check that the handle of the arm system locks securely into the transport safeguard.
2.2.6

Handle

SIM
Mounting
Check that the handle on the support arm is securely mounted. If necessary, tighten the
screws with a torque of 15-20 Nm.
2.2.7

Collimator adjustment knobs

SIM
Mounting
Check whether the adjustment knobs on the collimator are tightened securely
(Fig. 9 / p.
Siemens AG
Medical Solutions
31). If necessary, tighten both Allen screws on each of the knobs.
SPR8-220.101.04.03.02
07.05
Fig. 9:
(Fig. 8 / p. 31)
for mechanical damage. In case of dam-
Page 31 of 64
CS PS 24
31
MOBILETT Plus HP