Inspection and Maintenance
2.15

Checking the radiation indicator

Fig. 35:
Pos. 1
Set 81 kV and 10 mAs.
X
Release an exposure and check whether an indicator lights up on the control and dis-
play panel, and an audible signal sounds at the end of the exposure.
SIE
Radiation indicator
SIE
Audible signal
Siemens AG
Medical Solutions
Exposure light indicator
SPR8-220.101.04.03.02
07.05