Coincidence Of Radiation And Light Fields - Siemens MOBILETT Plus HP Maintenance Instruction

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

Inspection and Maintenance
2.14

Coincidence of radiation and light fields

Fig. 33:
Fig. 34:
NOTE
Insert a film into the 35 x 35 cm (14" x 14") cassette.
Place the cassette on a table or similar surface.
Set the SID to 100 cm (39") (use a tape measure).
Switch the collimator lamp on and align the cassette. Now collimate the light field to 25
x 25 cm (10" x 10").
Mark the four corners of the light field with coins and identify one of the corners with an
additional coin.
Set 52 kV and 2.0 mAs on the control and display panel.
X
Release an exposure.
Develop the film and measure the deviation between the light and radiation fields. A de-
viation of ± 1.7 % of the SID is acceptable.
If the deviation is larger than 1.7%, correct the coincidence of the light and radiation
fields by adjusting the nuts marked "D" and "E".
X
Check the setting by releasing another exposure.
Siemens AG
Medical Solutions
If the MOBILETT is equipped with a DAP measuring system, re-
move the DAP ionization chamber first.
(Fig. 33 / p. 55)
SPR8-220.101.04.03.02
07.05
.
(Fig. 34 / p. 55)
(Fig. 30 / p. 52)
Page 55 of 64
CS PS 24
.
MOBILETT Plus HP
55