Batteries

NOTE
PMF
Battery voltage
Check the battery voltage (200 V - 230 V) at K202 (16 each 12 V batteries in series).
PMF
Condition
In order to assess the condition of the batteries, ask the operating personnel whether
the batteries need recharging more often than previously. If this is the case, the batter-
ies should be replaced. See section "Replacement of batteries B10" in the Service In-
structions.
PMF
Charging indicator
Check that the charging indicator is functioning correctly.
Reattach all cover parts with the exception of the upper system cover.
MOBILETT Plus HP
Normal lifetime: 1 to 2 years.