Testing the siren

Smart-Vue Wireless Siren Installation Guide
13
This section describes how to run a sample test after adding the siren to your
system as described in the previous sections. The test procedure described
here generates a false alert with a time/date stamp of 12/30/1899 at 00:00 so
you can identify these events as tests easily. Also, they are not recorded in the
alarm log.
To test your siren, along with all other alerts you may have configured (fax,
SMS, e-mail, etc.), do the following. Enter your login and password when
prompted.
1. In Smart-Vue Client, select the Tools tab and then click on Alert test
(
).
2. Enter your login name and password to continue.
3. In the Alert test window, select the receiver or sensor you wish to
check. Groups (such as "Group 1" and "Lab1" in Figure 7) are not
tested.
Figure 7. Select the sensor and type of test