4. From the Select remote to be tested pick-list to choose the type of
alarm you wish to test (e.g., "High limit alarm" set in Error! Not a
valid link.
5. Click on Test ==> Close.
6. Click on the Alarms label in the lower left-hand corner of the main
screen to view the alarms zone. Within a few seconds, the test alarm
will appear in the alarms zone.
Figure 8. Select the sensor and type of test
The siren, as well as all other alerts configured for the receiver and/or sensor
in question, turn on within a few moments. This is a good way to test the
entire alert process in your wireless monitoring system.
7. Click on the red siren icon (
8. Click on the red exclamation point icon (
alarm and clear it from the alarm list.
9. Complete the fields (e.g., cause of incident, corrective action) on the
Alarm Acknowledgement screen and click on OK.
Software configuration
) to turn off the alarm.
) to acknowledge the
Smart-Vue Wireless Siren Installation Guide