Test Record Form

Product Serial Number:
Test Date:
Test Model:
T3HVD1500-70
DC Accuracy
T3HVD1500-200
T3HVD7000-100
T3HVD1500-70
Rise Time
T3HVD1500-200
T3HVD7000-100
T3HVD1500-70
DC Common
Mode Rejection
T3HVD1500-200
Ratio
T3HVD7000-100

Care and Maintenance

Keep the probe clean and dry.
To clean wipe with a soft dry cloth. Do not use chemicals or liquids.
Keep the probe in the case provided when not in use. This will help to keep the probe clean.
Do not drop the probe since this can cause shock and damage the probe.
Do not use force, excessively twist, excessively bend or pull on the input cables. This can cause damage
to the input cables and reduce the life of the probe.

Reference

IEC/EN 61010-031:2015 Definitions
Measurement Category III (CAT III) applies to test and measuring circuits connected to the distribution part
of the building's low-voltage mains installation.
Measurement Category II (CAT II) applies to test and measuring circuits connected directly to utilization
points (socket outlets and similar points) of the low-voltage mains installation.
No Rated Measurement Category applies to other circuits that are not directly connected to the mains
supply.
Pollution Degree 2 refers to an operating environment where normally only dry nonconductive pollution
occurs. Conductivity caused by temporary condensation should be expected.
Note: The 2015 version of the standard eliminates the Measurement Category I (CAT I) rating for test and
measuring circuits that are not intended to be directly connected to the mains supply. However, this change
in rating designation does not reduce the measurement capability or the level of protection offered by the
new probe design when compared to the probes that were originally certified per the 2008 version of the
standard.
Page 13
Test Temperature:
Test Humidity:
Lower Limit
50 x
98 mV
500 x
98 mV
50 x
98 mV
500 x
98 mV
100 x
98 mV
1000 x
98 mV
50 x
-
500 x
-
50 x
-
500 x
-
100 x
-
1000 x
-
50 x
-
50 x
-
100 x
-
Test result
Upper Limit
102 mV
102 mV
102 mV
102 mV
102 mV
102 mV
5 ns
5 ns
1.75 ns
1.75 ns
3.5 ns
3.5 ns
1 mV
1 mV
1 mV