4. The Object Dictionary
4.1.11

Object 1015h: Inhibit Time Emergency

The inhibit time emergency defines the inhibit time for the EMCY Message.
Index
Parameter Name
Object Type
Data Type
Access
Default Value
PDO Mapping
4.1.12

Object 1017h: Producer Heartbeat Time

The producer heartbeat time defines the cycle time of the heartbeat. If the time is 0 it is
not used.
The time has to be a multiple of 1 millisecond.
Index
Parameter Name
Object Type
Data Type
Access
Default Value
PDO Mapping
4.1.13

Object 1018h: Identity Object

This object contains general information about the device.
Sub-index 1: CANopen Vendor ID
Sub-index 2: Manufacturer specific product code
Sub-index 3: Revision number. Bit 31-16 is the major, Bit 15-0 the minor revision
number.
Sub-index 4: Manufacturer specific serial number.
Index
Parameter Name
Object Type
WIKA special documentation, model MH-4-CAN
1015h
Inhibit Time Emergency
VAR
UNSIGNED16
rw
0
No
1017h
Producer Heartbeat Time
VAR
UNSIGNED16
rw
0
No
1018h
Identity Object
RECORD
EN