Measurement group items when the Manual Mode Measurement feature is enabled.
Measure With
Gate A
Gate B
TCG
Menu Group - Limits
Limits
Maximum
Reference
Minimum
Menu Group - Probe
Auto Detect
Type
Zero
Autoset Gain
Menu Group - Calibration
1 Point Cal
2 Point Cal
Velocity
Material
Cygnus 1 Ex Operating Manual
Selects which Gate/Gates are used for thickness
measurements.
GateA SE = SE measurement using Gate A
GateA EE = EE measurement using Gate A
GateAB EE = EE measurement using Gate A & B
Gate A setup – Start, Width, Level, Flank/Peak/First-Peak.
Gate B setup – Start, Width, Level, Flank/Peak/First-Peak.
Time Controlled Gain settings – Gain, Slope, Start.
Turn the Limits feature On or Off.
Set the Maximum thickness value, or set to zero when not
required.
Set the Reference thickness value.
Set the Minimum thickness value.
Turn the Auto Probe Detect feature On or Off.
Select the Probe Type, if Auto Detect is off or the probe
can't be detected.
Start a Probe Zero measurement.
Start an Autoset Probe Gain measurement.
Start a Single Point Calibration.
Start a Two Point Calibration.
Set the Velocity of Sound value.
Set the velocity of sound using a Material from a list of
materials.
The material list is stored as a plain text file in the gauge's
User directory and can be modified to provide custom
material lists.