Advertisement

4.
Access the Menu and scroll down to
the 'Calibration' group. Then scroll
right to the '2 Point Cal' item.
Press the √ key to start.
5.
Start with the thick sample – the
maximum thickness.
6.
While holding the probe firmly on the
thick sample and while a steady
thickness measurement is displayed.
Use the up and down ▲▼ navigation
keys to adjust the thickness to the
required value.
7.
When the correct measurement is
displayed press the Set function key
A dialog box is displayed where you
can confirm the required thickness
measurement. Press the √ key to
accept or X key to cancel.
8.
Next measure the thin sample – the
minimum thickness.
Cygnus 1 Ex Operating Manual
Page 45 of 91