Diagnostics

Troubleshooting

MANAGING ERRORS
If during a sterilization cycle an error occurs do the following:
Wait until the OPEN button appears.
1
At the end of the reset phase, you may open the door. A pop
2
appears requiring a confirmation.
Tap
to open the door.
3
CAUTION! Do not switch off the sterilizer during the reset
phase: it takes some minutes to reset the system and
reach safe conditions in the sterilizer chamber.
Notice: water could be present in the chamber when opening the
door: prevent spilling (e.g., place a towel below the chamber door).
VIEW AND SAVE THE ERROR LOG
On the homepage, tap
1
appears.
Tap
to save the list in the USB pen drive.
2
74
>
>
: a list of the last errors
ERROR PAGE
During the sterilization cycle, the sterilizer is continuously monitored
by a control system. If an anomaly is detected, the cycle is aborted
automatically, and the sterilizer starts a reset phase.
The following page appears:
Lexa  | Instructions for Use |  MN-111 ENG Rev14 |  03/04/2023 |  © W&H Sterilization Srl
Part
Description
A
Current sterilization cycle
B
Error number, See "Errors"
on page  69 .
C
Warning messages.
D
Open button that appears
after the reset phase had
finished.