400-cycle maintenance

REPLACING THE HEPA FILTER
Notice: The HEPA filter needs to be replaced every 400 sterilization cycles or once a year, whichever come first. A replacement message alerts
when replacement is due. If the consumables is replaced prior to the message, you have to reset the consumable cycle counter.
Proceed as follows:
 
Open the sterilizer door.
1
 
Unscrew the HEPA filter by hand (counter-clockwise).
2
 
Screw on the new HEPA filter (clockwise) and tighten it snug.
3
Lexa PLUS  | Instructions for Use |  RIS-311 ENG Rev03 |  29/08/2022 |  © W&H Sterilization Srl
Maintenance
69