CLEANING AND DISINFECTION

5

Nebuliser handset

WARNING:
The nebuliser handset (and the aerosol
head) must be cleaned immediately after
each use and disinfected once a day.

Preparation

Disassemble the nebuliser handset into its
individual parts:
• Pull the connection cord out of the
nebuliser handset.
• Remove the mouthpiece from the
nebuliser handset.
• Open the medication reservoir and pour
out the medication residue.
Info:
The medication reservoir and aerosol
chamber are already preassembled
and must not be detached from one
another.
• Open the aerosol chamber (6a).
®
eFlow
rapid - 2022-07
• Remove the inspiratory valve from the
aerosol chamber.
• Remove the aerosol head (4).
To do this, gently press the engaging
side hooks on the aerosol head
together.
CAUTION:
Do not touch the membrane (the
curved metal surface in the middle of
the aerosol head) to avoid damage.
6a
en
4
17