Fault/Condition
The controller cannot be switched on
(no beep, no orange or green light
signal).
After switching the controller on, no
aerosol is generated, and/or the
controller switches off again after a
few seconds.
The controller switches off even
though there is still medication left in
the medication reservoir.
Info:
Note that about 1 ml of
medication cannot be
nebulised and remains unused
in the medication reservoir of
®
the eFlow
rapid nebuliser
handset; this residue must be
discarded. This is intentional
and does not represent a
malfunction.
With drug-specific nebuliser
handsets (Tolero
®
Altera
) there is no significant
volume of residue left in the
medication reservoir.
The controller does not switch off
automatically even when no more
aerosol is being produced and there
is no more than about 1 ml of residue
left in the medication reservoir of the
®
eFlow
rapid nebuliser handset.
®
eFlow
rapid - 2022-07
Possible cause/Solution
- To switch on, press the
- Is power cord plug plugged into mains
- Are the batteries inserted correctly?
- Check the charge level of the batteries.
- Has the medication reservoir been filled
- Check the connection between the
If there is still more than 1 ml in the
medication reservoir of the eFlow
nebuliser handset:
- Check the charge level of the batteries.
- Was the nebuliser handset held
- The controller switches off
®
®
, Zirela
and
The treatment session has been
completed successfully and the controller
can be switched off by pressing the
ON/OFF button.
ON/OFF button for approx. 2 seconds.
and the device (for mains operation)?
Does the power adapter LED light up?
with medication?
controller and the nebuliser handset.
horizontally throughout the inhalation
session? Hold the nebuliser handset
horizontally.
automatically after 20 minutes.
Carry out inhalation to the end by
pressing the ON/OFF button again, so
that you can inhale the prescribed dose
of medication.
If the maximum operating time for an
inhalation session is exceeded
repeatedly, the aerosol head must be
replaced.
en
®
rapid
23