TROUBLESHOOTING

6
You can correct malfunctions that occur during daily operation yourself. The following
tables will help you to locate and correct the causes of malfunction.
Screen
(flashes alternately)
22
Possible cause
Connection
between the
controller and the
nebuliser handset/
aerosol head
interrupted.
No medication
added.
Inhalation
interrupted.
The maximum
operating time of
20 minutes for an
inhalation session
has been
exceeded.
Battery charge
level 50%.
Battery flat.
Solution
- Verify that all parts
are securely
connected.
- Replace the aerosol
head and/or the
connection cord if
damaged.
If the fault persists,
contact your service
partner.
Add medication.
Resume inhalation by
restarting the device
(press the
ON/OFF button).
Resume inhalation by
restarting the device
(press the
ON/OFF button).
Have new batteries
ready.
Insert new batteries or
operate with the power
adapter connected.
®
eFlow
rapid - 2022-07