• Place the individual parts on a dry, clean
surface and allow them to air-dry
completely.
CAUTION:
Moisture promotes the growth of
bacteria. Please ensure adequate
drying after every disinfection.
CAUTION:
Check the nebuliser handset parts after
every disinfection. Replace all broken,
deformed and heavily discoloured parts.
Hygienic reprocessing in the
hospital and doctor's
practice
Carry out cleaning with disinfection
immediately after use. Ideally, a mechani-
cal procedure (instrument dishwasher)
should be used.
Prepare for cleaning with disinfection
(see "CLEANING AND DISINFECTION",
page 17).
20

Cleaning with thermal disinfection

• Place the nebuliser handset compo-
nents in the instrument dishwasher.
• Select the 93 °C programme (treatment
time: 10 minutes).
The effectiveness of this procedure has
been validated using a G7736 disinfector
made by Miele and using neodisher MA
(Chemische Fabrik Dr. Weigert,
Hamburg) as a cleaning agent in conjunc-
tion with deionised water as a neutraliser.
Even when the instrument dishwasher is
equipped with a drying function, ensure
that no residual moisture remains in the
nebuliser handset. Remove residual
moisture by shaking the parts and allow
them to air-dry completely.
®