Confirming Speed Specification; Tbl. 14: Characteristic Nominal Rotation Speeds Of The Turbopumps; Fig. 23: Installation Of Electronic Drive Unit Tc 400 - Pfeiffer Vacuum HiPace 800 Operating Instructions Manual

Turbopump
Table of Contents

Advertisement

1
Fig. 23:
1 Hexagon socket screw
2 Electronic drive unit
3 Adapter plate
Procedure
1. Place a new electronic drive unit straight onto the adapter plate connection of the turbopump.
2. Screw the electronic drive unit to the turbopump with all 3 Allen head screws.
7.4.3 Confirming speed specification
The typical nominal rotation speed of a turbopump is preset ex factory in the electronic drive unit. If the
electronic drive unit is replaced or a different pump type is used, the set value settings of the nominal
rotation speed is cleared. The manual confirmation of the nominal rotation speed is part of a redundant
safety system as a measure for preventing excess rotation speed.
The redundant confirmation of the nominal rotation speed of a turbopump is possible by adjusting the
[P:777] NomSpdConf parameter in the electronic drive unit.
HiPace
10 | 30 | 60 | 80
300
350 | 450
400 | 700 | 800
Tbl. 14:
Required aids
● Connected Pfeiffer Vacuum control unit
● Knowledge of the configuration and setting of electronic drive unit operating parameters
Adjusting nominal rotation speed confirmation
1. Observe the operating instructions of the control unit.
2. See the electronic drive unit operating instructions.
3. Set the parameter [P:794] to "1" and activate the expanded parameter set.
4. Open and edit the parameter [P:777].
5. Set the parameter [P:777] to the required value of the nominal rotation speed in Hertz.
3
2
Installation of electronic drive unit TC 400
– Tightening torque: 2.5 Nm
Characteristic nominal rotation speeds of the turbopumps
Alternative to adjusting the nominal rotation speed confirmation
A Pfeiffer Vacuum SpeedConfigurator for the one-time immediate setting of parameter
[P:777] is included with the replacement units.
4
High vacuum connection protective cover
5
Fore-vacuum connection protective cover
Nominal rotation speed
1500 Hz
1000 Hz
1100 Hz
820 Hz
4
5
Maintenance
45/64

Advertisement

Table of Contents