Installation
Fig. 16:
1 Turbopump fore-vacuum connection
2 Centering ring
3 Circlip
Establish the fore-vacuum connection
1. With rigid pipe connections, include bellows to attenuate external vibrations.
2. Install a fore-vacuum connection with small flange components, e.g. connection elements and
3. Implement measures to counteract the backflow of operating fluids or condensate from the fore-
4. Observe the information in the operating instructions of the backup pump or pumping station when
5.5 Cooling water connection
Risk of scalding from suddenly escaping hot cooling water
The turbopump water connections are open to both sides. When connecting the cooling water supply,
there is a risk of scalding from suddenly escaping, hot cooling water at over pressure.
► Prior to installation, ensure that pressure is discharged from the cooling water system, and that it
► Wear protective equipment, e.g., safety goggles and gloves.
30/64
Suitable backing pump
Use the turbopump only in combination with a suitable backing pump that can deliver up to
the required maximum fore-vacuum pressure. To achieve the fore-vacuum pressure, use a
suitable vacuum pump or a pumping station from the Pfeiffer Vacuum range.
In this case, the backing pump is also controlled directly via the turbopump electronic drive
unit interfaces (e.g., relay box or connection cable).
Example of fore-vacuum connection on HiPace 800
pipe components DN 25 ISO-KF from the
vacuum area.
connecting and operating it.
WARNING
is cooled down.
2
1
3
4
Vacuum components DN 25 ISO-KF
5
Backing pump (e.g. multi-stage roots pump)
Pfeiffer Vacuum Components
4
5
Shop.