Foil Clamp
The foil clamp is used for foils or thin specimens.
To insert the specimen, loosen the three clamping screws [a] slightly and push the movable jaw [c]
sideways against the two springs [b].
The specimen is kept in place by the springs [b]; however, its posi- tion can be changed. Turn the three
clamping screws [a] to clamp the specimen tightly against the fixed jaw [d].
When using an orienting adapter with the instrument, first insert the enclosed graduated ring into the
orienting adapter with the pin. With the graduated ring the orientation in X-/Y-direction is can- celled.
However, it is possible to turn the Z-axis 60° in either direction.
Then insert the foil clamp.
Note
According to the various specimens, it might be helpful to use in addition Thermo Fisher Scientific's sandwich
supporting material (Cat. No. 176010).