Clearance Angle Adjustment
The clearance angle between cutting edge and specimen can be shifted and adjusted to the requirements of the
tissue to be sectioned.
Loosen the clamping lever [h] on the right side of the blade carrier and move the upper part [k] of the
blade carrier on the base [a].
The adjusted clearance angle can be read on the scale of the upper part [k].
Then turn the clamping lever [h] upwards to lock in the new clea- rance angle.
Note
By experience, usable cuts are only achieved at a clearance angle of 10° or more.
Moving the Blade Carrier on the Console
If the clamping lever [h] is loosened, the upper part [k] of the blade carrier can additionally be moved 1 cm to
the left or right side. This way, the cutting edge can optimally be used. The clamping lever for the angle
adjustment can be pulled off after a correct angle setting to avoid that the angle is shifted accidentally.
Loosen the clamping lever [f ] on the left side of the blade carrier, to move the carrier forwards or
backwards on the guide bars.This allows a rough adjustment of the blade to the specimen.
Protection against Injury
A knife guard [d] on the clamping plate can be moved upwards over the blade for the protection against injury.