Bender COMTRAXX CP9 Series Manual page 35

Remote alarm indicator and operator panel for medical locations and other areas
Hide thumbs Also See for COMTRAXX CP9 Series:
Table of Contents

Advertisement

COMTRAXX® CP9xx – Control Panel
Modbus
Bender Modbus image
V1, V2 (V2)*
Modbus TCP
Interface/protocol
Ethernet/Modbus TCP
Cable length
< 100 m
Operating mode
client for Bender Modbus TCP devices and 'third-party
devices'
Operating mode
Server for access to process image and for Modbus
control commands
Parallel data access from different clients
max. 25
Modbus RTU
Interface/protocol
RS-485/Modbus RTU
Cable length
< 1200 m
Cable
shielded, one end of shield connected to PE
recommended
CAT6/CAT7 min. AWG23
alternative
twisted pair, J-Y (St) Y min. 2x0.8
Connection
'AMB', 'BMB' (see plug-in terminal)
Operating mode
master/slave (master)*
Baud rate
9.6...57.6 kBit/s
Terminating resistor
120 (0.25 W), can be connected internally (see plug-in
terminal)
Supported Modbus RTU slave addresses
2...247
PROFINET
Interface/protocol
Ethernet/PROFINET
Operating mode
slave (IO device)
SNMP
Interface/protocol
Ethernet/SNMP
Versions
1, 2c, 3
Supported devices
query of all devices (channels) possible
Trap support
no
CP9xx_D00349_09_M_XXEN/03.2023
35