Connect terminals A1/+ and A2/- (4) to the power source. Connect PE to earth. The CP907-I can also
be supplied via Power-over-Ethernet (PoE). The PoE switch must be earthed. For further details, see
technical data.
8.
Attach the front plate to the built-in ush-mounting enclosure.
CP907 is mounted. The following steps apply to CP915 or CP924 only:
9.
Connect the front panel to the control board and the power supply:
Connect a USB socket (8) to the touch sensor connector on the front panel. Associated connection cable
included in the scope of delivery.
10. Connect the screen output DVI (9) to the front plate. DVI cable: Cable length < 3 m, connect rmly to PE
on both sides.
11. When using the audio output (10), an electrically isolated ampli er must be used.
12. Connect the voltage supply to the power supply unit via the pre-assembled wiring. Connect the earthing
to the front plate.
Pre-assembled voltage supply
Connection earthing on front plate