Bender COMTRAXX CP9 Series Manual page 17

Remote alarm indicator and operator panel for medical locations and other areas
Hide thumbs Also See for COMTRAXX CP9 Series:
Table of Contents

Advertisement

Remove the CP9xx from the built-in ush-mounting enclosure.
Mainboard and connections of the CP9xx device
No.
1
Plug connector digital inputs
2
I²C interface
3
Plug connector to energy storage board
4
Voltage supply A1/+, A2/–, PE
5
Ethernet (RJ45/CAT5); HTTP, Modbus TCP, BCOM
6
X1 plug connector for Modbus RTU, BMS bus
7
Termination of Modbus RTU and BMS bus
8
USB ports (for touch sensor)
9
DVI output
10
Audio output
11
Audio input
12
Connection to control relay
Connect the device as follows:
1.
Modbus RTU connection (6): Connect terminals AMB and BMB to the Modbus RTU (A to A, B to B).
2.
BMS bus connection (6): Connect terminals ABMS and BBMS to the BMS bus (A to A, B to B)
X1 plug assignment (6)
Connection
COMTRAXX® CP9xx – Control Panel
CP907
with PoE
not included
not included
not included
not included
CP9xx_D00349_09_M_XXEN/03.2023
CP915/CP924
without PoE
17