Bender COMTRAXX CP9 Series Manual page 15

Remote alarm indicator and operator panel for medical locations and other areas
Hide thumbs Also See for COMTRAXX CP9 Series:
Table of Contents

Advertisement

For surface mounting, the ush-mounting enclosure is mounted in the optionally available surface-mounting
enclosure (B95061915).
Enclosure for surface mounting CP907, dimensions in mm
Mounting procedure:
1.
Assemble surface-mounting enclosure ( t hinges and bracket).
2.
Insert the ush-mounting enclosure through the opening in the cover. Mount the enclosed plastic frame
from behind and screw it in place using the fasteners of the ush-mounting enclosure.
3.
Make the required cable openings in both enclosures.
4.
Connect the CP907 and mount it in the enclosure.
Mounting CP915/CP924
Flush-mounting enclosure CP915/CP924
Mounting of ush-mounting enclosures
The supplied ush-mounting enclosures are only suitable for mounting in cavity walls. In drywall and
stud frame constructions, the enclosures must be screwed at right angles to battens or to the stud frame
structure. The enclosure must not warp during mounting. The wall surface must be even.
Flush-mounting enclosure CP915/CP924, dimensions in mm
COMTRAXX® CP9xx – Control Panel
CP9xx_D00349_09_M_XXEN/03.2023
15