Bender COMTRAXX CP9 Series Manual page 23

Remote alarm indicator and operator panel for medical locations and other areas
Hide thumbs Also See for COMTRAXX CP9 Series:
Table of Contents

Advertisement

7.
If there is a DHCP server in the network, select only the check box to the right of the 'DHCP?' label in line
4. Con rm your selection by pressing the 'Save' button. The network settings transmitted from the server
are shown on the display after 8...10 seconds.
After this, the commissioning must be continued via the COMTRAXX® user interface.
Factory settings communication addresses
IP address
IP address for 1:1 ETH conn.
Net mask
Standard gateway
DNS
DHCP
t
Timeout for address assignment
off
BMS address
BMS protocol
BCOM system name
Subsystem address
BCOM device address
The settings can be changed during commissioning via the display or the web user interface.
• From an external application (e.g. visualisation software), commands can be sent to BMS devices. The
'Modbus control commands' menu provides Modbus control commands for selected BMS commands.
These commands can be copied to the clipboard of the PC and then included in the programming of the
external application.
• Graphical display with scaling of the time axis for the data loggers of the gateway and compatible Bender
devices.
BMS interface
The majority of Bender devices communicate via the internal BMS bus. CP9xx can be operated as a master or
as a slave.
CP9xx is to be operated as a master if:
• Parameters are queried or changed
• Certain control commands are issued
Note that not all BMS masters can surrender their master function!
Parameter
COMTRAXX® CP9xx – Control Panel
CP9xx_D00349_09_M_XXEN/03.2023
Factory setting
-
169.254.0.1
255.255.0.0
192.168.0.1
-
aus
30 s
1
BMS i
SYSTEM
1
1
23