Digital Inputs - Bender COMTRAXX CP9 Series Manual

Remote alarm indicator and operator panel for medical locations and other areas
Hide thumbs Also See for COMTRAXX CP9 Series:
Table of Contents

Advertisement

4.2

Digital inputs

CP9xx-devices feature 12 con gurable digital inputs. Settings are made via the COMTRAXX® user interface in
a browser.
Function
The following functions can be assigned to the digital inputs:
o
Digital input without function
TEST
Self test of the device
RESET
Reset fault and alarm messages
Menu > Settings > Digital input 1 - 12
For each of the 12 digital inputs DI1...12, the following can be de ned:
Parameter
Mode
Measurement type
t
on
t
off
An event is executed when the digital input experiences an edge change. The edge change must be present
at least for the set response delay t
High-active
Low-active
Impuls (High-active)
Impuls (Low-active)
Operating message
Alarm
Error(s)
Response delay
Switch-off delay
, otherwise it is ignored.
on
COMTRAXX® CP9xx – Control Panel
Options/Setting range
CP9xx_D00349_09_M_XXEN/03.2023
0 s...10 minutes
21