Bender COMTRAXX CP9 Series Manual page 16

Remote alarm indicator and operator panel for medical locations and other areas
Hide thumbs Also See for COMTRAXX CP9 Series:
Table of Contents

Advertisement

Removing the CP915/CP924 front plate
The front plate is removed from the enclosures of the CP915 and CP924 devices using a suction lifter. For this
purpose, the suction lifter must be placed on the points marked below one after the other and the front plate
must be removed until it clicks into place for the rst time. If the front plate is detached on both sides, the
plate can be lifted o the enclosure.
CAUTION
Placing the suction lifter in the middle and pulling only at this point may damage the front plate.
Always place the suction lifter on the edges of the display.
Connection of the CP9xx device
The CP9xx is integrated into existing LAN structures, but can also be operated via a single PC.
Con guration of computer networks
If you are familiar with the con guration of computer networks, you can carry out the connection of the
device yourself. Otherwise please contact your IT administrator!
Ethernet
The shield of the Ethernet cable must be connected to PE on both sides.
For UL applications (CP907 only)
Use copper lines only. Minimum temperature range of the wires to be connected to the plug-in terminals:
75 °C.
PoE (CP907 only)
Minimum temperature range of the cables (copper lines) to be connected to the PoE Ethernet connection:
80 °C
For operation via PoE, the voltage transmitter (router) must meet one of the following requirements:
- Class 2 requirement acc. to UL1310 or
- Limited power source requirement acc. to UL 60950 or
- Limited energy circuit requirement acc. to UL 61010.
In case of a pure PoE supply, no supply of the I²C expansion modules is possible. Maximum cable length
I²C < 3 m
16
CP9xx_D00349_09_M_XXEN/03.2023