Atmos C 341 Battery Operating Instructions Manual page 59

Hide thumbs Also See for C 341 Battery:
Table of Contents

Advertisement

Charging time
Recharging interval during long-term
storage
Battery operating time during
continuous operation with fully
recharged battery / new battery (>20 l/
min, setting -0.8 bar)
Life cycle
Display
Typical battery operating life*
* Measured at +21° C, continuous operation, without battery recharging and at free air
flow.
DDS canister system
Capacity
Connection reusable suction hose
Reusable suction hose:
Diameter
Length
Connection to the suction device
Hydrophobic DDS bacterial and viral
filter
Degree of separation against bacteria
Degree of separation against viruses
Total degree of separation
Filter class
*External test report (test laboratory)
Single-use canister systems
Capacity
Connection disposable suction hose
Disposable suction hose:
Diameter
Length
Connection to the suction device
Bacterial filter
Battery status 80%: 2 h 40 min
Battery status 100%: 14 h
Automatic switch over to trickle charging
Every 3 months.
23 min at -5° C
40 min at +21° C
40 min at +50° C
400 recharging cycles within approx. 3 years
Display of battery status during operation
and recharging
-0.2 bar: 40 min
-0.5 bar: 40 min
-0.8 bar: 40 min
1000 ml
Ø 10 mm I.D.
Ø 10 mm I.D.
1300 mm
Direct connection (without intermediate
hose)
Hydrophobic DDS bacterial and viral filter
cartridge for use in the secretion canister lid,
single-use product
99.999778%*
99.73%*
>99.95%*
H13 (High-Efficiency Particulate Air/
Arrestance)*
1000 ml
Ø 7 mm I.D.
Ø 6 mm I.D.
1300 mm
By means of a vacuum hose (intermediate
hose)
Integrated in the suction bag
Technical data
59