DDS canister system
Number of reprocessing cycles: max. 50.
1. Disassemble canister lid and filter holder into their individual parts. Chapter „3.4
Connection and removal of canister system and hoses" on page 21.
2. Rinse the following parts of the DDS canister system with clear water:
-
-
-
-
-
-
-
-
3. Clean the mentioned parts with a brush or a cloth.
4. Disinfect the mentioned parts with a disinfectant which is recommended by ATMOS.
5. Let the individual parts of the canister lid and the filter holder dry.
As soon as the individual parts are dry:
6. Insert a new hydrophobic DDS bacterial and viral filter.
7. Reassamble the individual parts of the canister lid and the filter holder.
Serres
Please observe the instructions in the operating instructions for the canister system.
Do not operate the device without a bacterial filter. The hydrophobic bacterial and
viral filter (REF 443.0738.0) must be used with a secretion canister system which has
no integrated bacterial filter.
Vacuum hose
After every suction process:
1. Rinse the vacuum hose with clear water for at least 10 seconds.
It must be exchanged after each patient or at least once a day:
2. Disinfect the vacuum hose with a disinfectant for the accessories recommended by
ATMOS.
38
Cleaning and disinfection
1
2
Filter holder
1
Sealing ring
2
Secretion canister
Inner canister lid
Outer canister lid
Float ball
Filter holder
Sealing ring
Suction hose
Support for canister system
, Medi-Vac
secretion canister system, other canister systems
®
®
3
4
Inner canister lid
3
Float ball
4
Outer canister lid
5
5