After Cleaning - Atmos C 341 Battery Operating Instructions Manual

Hide thumbs Also See for C 341 Battery:
Table of Contents

Advertisement

Device surface
WARNING
Electric shock by liquid in the device.
Disconnect the device from the mains power supply prior to cleaning.
Do not rinse the device under running water and do not immerse it into any liquids.
Make sure that the cleaning cloth is only damp and not wet.
Do not autoclave the device.
Do not immerse the device in disinfectant solution.
1. Clean the entire surface of the device and the hose rewind with a damp cloth.
2. Disinfect the entire surface of the device and the hose rewind with a surface
disinfectant.
Power supply and recharging unit
WARNING
Electric shock by liquid in the power supply.
Disconnect the power supply and recharging unit from the mains power supply
prior to cleaning.
Do not rinse the power supply and recharging unit under running water and do not
immerse it into any liquids.
Make sure that the cleaning cloth is only damp and not wet.
Do not autoclave the power supply and recharging unit, do not sterilize it and do
not thermally disinfect it.
Do not immerse the power supply and recharging unit in disinfectant solution.
1. Clean the power supply and recharging unit with a damp cloth. You can use a mild
detergent.
2. Disinfect the power supply and recharging unit with a surface disinfectant.
Recommended: Terralin
Wall and device support
1. Clean the wall and device support with a damp cloth.
2. Disinfect the wall and device support with a surface disinfectant.

5.3 After cleaning

WARNING
Risk of injury to user and patient due to a damaged device.
1. Check after each cleaning, whether the device is obviously damaged. In case the
device is damaged, please send it in to ATMOS.
2. Perform a function check: Chapter „7.2 Functional check" on page 47.
3. Prepare the device for the next use.
Protect
®
Cleaning and disinfection
39